Agilent Technologies E8247C PSG CW User Manual

Page 83

Advertising
background image

Chapter 3

73

Optimizing Performance

Creating and Applying User Flatness Correction

2. Configure the signal generator to interface with the power meter.

a. Press

Amplitude

>

More (1 of 2)

>

User Flatness

>

More (1 of 2)

>

Power Meter

>

E4416A

,

E4417A

,

E4418B

,

or

E4419B

.

b. Press

Meter Address

> enter the power meter’s GPIB address >

Enter

.

c. For E4417A and E4419B models, press

Meter Channel A B

to select the power meter’s active channel.

d. Press

Meter Timeout

to adjust the length of time before the instrument generates a timeout error if

unsuccessfully attempting to communicate with the power meter.

3. Press

More (2 of 2)

>

Configure Cal Array

>

More (1 of 2)

>

Preset List

>

Confirm Preset

.

This opens the User Flatness table editor and resets the cal array frequency/correction list.

4. Press

Configure Step Array

.

This opens a menu for entering the user flatness step array data.

5. Press

Freq Start

>

26.5

>

GHz

.

6. Press

Freq Stop

>

40

>

GHz

.

7. Press

# of Points

>

28

>

Enter

.

This enters the desired flatness-corrected frequencies (26.5 GHz to 40 GHz in 500 MHz intervals) into
the step array.

8. Press

Return

>

Load Cal Array From Step Array

>

Confirm Load From Step Data

.

This populates the user flatness correction array with the frequency settings defined in the step array.

9. Press

Amplitude

>

0

>

dBm

.

10. Press

RF On/Off

.

This activates the RF output and the

RF ON

annunciator is displayed on the signal generator.

Perform the User Flatness Correction

NOTE

If you are not using an Agilent E4416A/17A/18B/19B power meter, or if your power meter
does not have a GPIB interface, you can perform the user flatness correction manually. For
instructions, see

Performing the User Flatness Correction Manually

below.

1. Press

More (1 of 2)

>

User Flatness

>

Do Cal

.

This creates the user flatness amplitude correction value table entries. The signal generator begins the
user flatness correction routine and a progress bar is shown on the display.