Agilent Technologies E8247C PSG CW User Manual

Page 40

Advertising
background image

30

Chapter 2

Basic Operation
Configuring the RF Output

Setting the RF Output Amplitude

1. Preset the signal generator: Press

Preset

.

The

AMPLITUDE

area of the display shows the minimum power level of the signal generator. This is the

normal preset RF output amplitude.

2. Turn on the RF output: Press

RF On/Off

.

The display annunciator changes to

RF ON

. At the RF OUTPUT connector, the RF signal is output at the

minimum power level.

3. Change the amplitude to

20 dBm: Press

Amplitude

>

20

>

dBm

.

The new output power displays in the

AMPLITUDE

area of the display and in the active entry area. Until

you press a different front panel function key, amplitude remains the active function. You can also
change the amplitude using the up and down arrow keys or the knob.

Setting the Amplitude Reference and Amplitude Offset

The following procedure sets the RF output power as an amplitude reference to which all other amplitude
parameters are relative. The amplitude initially shown on the display is 0 dB (the power output by the
hardware minus the reference power). Although the display changes, the output power does not change. Any
subsequent power changes are shown as incremental or decremental to 0 dB.

1. Press

Preset

.

2. Set the amplitude to

20 dBm: Press

Amplitude

>

-20

>

dBm

.

3. Activate the amplitude reference mode and set the current output power (

20 dBm) as the reference

value: Press

More (1 of 2)

>

Ampl Ref Set

.

The

AMPLITUDE

area displays

0.00 dB

, which is the power output by the hardware

(−

20 dBm) minus

the reference value (

20 dBm). The

REF

indicator activates and the

Ampl Ref Off On

softkey toggles On.

4. Turn the RF output on: Press

RF On/Off

.

The display annunciator changes to

RF ON

. The power at the RF OUTPUT connector is

20 dBm.

5. Change the amplitude increment value to 10 dB: Press

Incr Set

>

10

>

dB

.

6. Use the up arrow key to increase the output power by 10 dB.

The

AMPLITUDE

area displays

10.00 dB

, which is the power output by the hardware

(-20 dBm plus 10 dBm) minus the reference power (

20 dBm). The power at the RF OUTPUT

connector changes to

10 dBm.

7. Enter a 10 dB offset: Press

Ampl Offset

>

10

>

dB

.

The

AMPLITUDE

area displays

20.00 dB

, which is the power output by the hardware (

10 dBm) minus

the reference power (

20 dBm) plus the offset (10 dB). The

OFFS

indicator activates. The power at the

RF OUTPUT connector is still

10 dBm.

Advertising