background image

Chapter 2

41

Basic Operation

Configuring the RF Output

4. Move the cursor back to marker 0 and press

Delta Ref Set

>

Marker Delta Off On

to On.

In the table editor, notice that the frequency values for each marker are now relative to marker 0.

Ref

appears in the far right column (also labeled

Ref

) to indicate which marker is the reference. Refer to

Figure 2-4

.

Figure 2-4

Marker Table Editor

5. Move the cursor back to marker 1 and press

Marker Freq

. Turn the front panel knob while observing

marker 1 on the 8757D.

On the 8757D, notice that the displayed amplitude and frequency values for marker 1 are relative to
marker 0 as the marker moves along the trace. Refer to