138

Chapter 6

Custom Arb Waveform Generator
Working with Modulation Types

4. Press

Return

>

Goto Row

>

0011 0000

>

Enter

; this is row 48.

5. Press the

Delete Row

softkey 16 times.

Repeat this pattern of steps using the following table:

6. Press

Display I/Q Map

to view the new constellation that has been

created. The I/Q State Map in this example has 128 symbols.

7. Press

Return

. When the contents of an

I/Q Values

table have not

been stored,

I/Q Values (UNSTORED)

appears on the display.

8. Press

More (1 of 2)

>

Load/Store

>

Store To File

.

If there is already a file name from the

Catalog of IQ Files

occupying the active entry area, press the following keys:

Editing Keys

>

Clear Text

9. Enter a file name (for example,

128QAM

) using the alpha keys and

the numeric keypad.

10. Press

Enter

.

The user-defined I/Q State Map should now be stored in the

Catalog of IQ Files

.

Goto Row

...

Press the

Delete Row

softkey...

0110 0000 (96)

16 times

1001 0000 (144)

16 times

1100 0000 (192)

16 times

0001 0000 (16)

4 times

0001 0100

(20)

4 times

0001 1000

(24)

8 times

0011 0000

(48)

4 times

0011 0100

(52)

4 times

0011 1000

(56)

4 times

0101 1000

(88)

8 times

0111 0000

(112)

4 times

0111 0100

(116)

4 times

0111 1000

(120)

8 times