174

Chapter 8

Multitone Waveform Generator
Creating, Viewing, and Optimizing Multitone Waveforms

6. Press

Edit Item

>

-10

>

dB

.

7. Highlight the value (

0

) in the

Phase

column for the tone in row 4.

8. Press

Edit Item

>

123

>

deg

.

9. Press

Apply Multitone

.

NOTE

Whenever a change is made to a setting while the multitone generator is operating (

Multitone

Off On

set to On), you must apply the change by pressing the Apply Multitone softkey before

the updated waveform will be generated. When you apply a change, the baseband generator
creates a multitone waveform using the new settings and replaces the existing waveform in
ARB memory.

You have now changed the number of tones to 10, disabled tone 2, and changed the power and phase of tone
4.

Figure 8-4 on page 174

shows what the multitone setup table display on the signal generator should look

like after all steps have been completed. The spectrum analyzer should display a waveform similar to the
one shown in

Figure 8-5 on page 175

. Notice that even-numbered multitone waveforms have a small amount

of carrier feedthrough at the center carrier frequency.