Figure 3-3 – Agilent Technologies E8247C PSG CW User Manual

Page 72

Advertising
background image

62

Chapter 3

Optimizing Performance
Using External Leveling

Determining the Leveled Output Power

Figure 3-3

shows the input power versus output voltage characteristics for typical Agilent Technologies

diode detectors. Using this chart, you can determine the leveled power at the diode detector input by
measuring the external detector output voltage. You must then add the coupling factor to determine the
leveled output power. The range of power adjustment is approximately -20 to +25 dBm.

Figure 3-3

Typical Diode Detector Response at 25° C

Advertising