Page 138

Advertising
background image

128

Chapter 6

Custom Arb Waveform Generator
Working with Filters

7. Press

Display Impulse Response

. A graph displays the impulse response of the current FIR coefficients.

8. Press

Return

.

9. Highlight coefficient 15.

10. Press

0

>

Enter

.

11. Press

Display Impulse Response

.

The graphic display can provide a useful troubleshooting tool (in this case, it indicates that a coefficient
value is set incorrectly, resulting in an improper Gaussian response).

12. Press

Return

.

13. Highlight coefficient 15.

14. Press

1

>

Enter

.

15. Press

Load/Store

>

Store To File

.

16. Name the file NEWFIR2, and press

Enter

.

The contents of the current FIR Values editor are stored to a file in the Memory Catalog and the Catalog
of FIR Files is updated to show the new file.