Agilent Technologies E8247C PSG CW User Manual

Page 78

Advertising
background image

68

Chapter 3

Optimizing Performance
Creating and Applying User Flatness Correction

Save the User Flatness Correction Data to the Memory Catalog

This process allows you to save the user flatness correction data as in the signal generator’s memory catalog.
With several user flatness correction files saved to the memory catalog, any file can be recalled, loaded into
the correction array, and applied to the RF output to satisfy specific RF output flatness requirements.

1. Press

Load/Store

.

2. Press

Store to File

.

3. Enter the file name

FLATCAL1

using the alphanumeric softkeys, numeric keypad, or the knob.

4. Press

Enter

.

The user flatness correction array file FLATCAL1 is now stored in the memory catalog as a

UFLT

file.

Applying a User Flatness Correction Array

Press

Return

>

Return

>

Flatness Off On

.

This applies the user flatness correction array to the RF output. The

UF

indicator is activated in the

AMPLITUDE

section of the signal generator’s display and the frequency correction data contained in the

correction array is applied to the RF output amplitude.

Recalling and Applying a User Flatness Correction Array

Before performing the steps in this section, complete

“Creating a User Flatness Correction Array” on

page 64

.

1. Press

Preset

.

2. Press

Amplitude

>

More (1 of 2)

>

User Flatness

>

Configure Cal Array

>

More (1 of 2)

>

Preset List

>

Confirm Preset

.

3. Press

More (2 of 2)

>

Load/Store

.

4. Ensure that the file

FLATCAL1

is highlighted.

5. Press

Load From Selected File

>

Confirm Load From File

.

This populates the user flatness correction array with the data contained in the file

FLATCAL1

. The user

flatness correction array title displays

User Flatness: FLATCAL1

.

6. Press

Return

>

Flatness Off On

.

This applies the user flatness correction data contained in

FLATCAL1.

Advertising