Agilent Technologies E8247C PSG CW User Manual

Page 53

Advertising
background image

Chapter 2

43

Basic Operation

Configuring the RF Output

3. Press

Sweep Time

to Manual >

5

>

sec

.

In auto mode, the sweep time automatically sets to the fastest allowable value. In manual mode, you can
select any sweep time slower than the fastest allowable. The fastest allowable sweep time is dependent
on the number of trace points and channels being used on the 8757D and the frequency span.

4. Press

Sweep Time

to Auto.

The sweep time returns to its fastest allowable setting.

Using Alternate Sweep

1. Press the

Save

hardkey.

This opens the table editor and softkey menu for saving instrument states. Notice that the

Select Reg

softkey is active. (For more information on saving instrument states refer to

“Using the Instrument State

Register” on page 54

.)

2. Turn the front panel knob until you find an available register and press

SAVE

. Remember this saved

register number. If no registers are available, you can write over an in-use register, by pressing

Re-SAVE

.

NOTE

When you are using the PSG in a system with an 8757 network analyzer, you are limited to

using registers 1 through 9 in sequence 0 for saving and recalling states.

3. Press

Sweep/List

>

Configure Ramp/Step Sweep

and enter new start and stop frequency values for the ramp

sweep.

4. Press

Alternate Sweep Register

and turn the front panel knob to select the register number of the

previously saved sweep state.

5. Press

Alternate Sweep Off On

to On.

The signal generator alternates between the original saved sweep and the current sweep. You may need
to adjust 8757D settings to effectively view both sweeps, such as setting channel 2 to measure sensor A.
Refer to

Figure 2-6

.