Figure 3-6 – Agilent Technologies E8247C PSG CW User Manual

Page 82

Advertising
background image

72

Chapter 3

Optimizing Performance
Creating and Applying User Flatness Correction

Figure 3-6

User Flatness with mm-Wave Source Module and Option 1EA Signal Generator

NOTE

To ensure adequate RF amplitude at the mm-wave source module RF input when using
Option 1EA signal generators, maximum amplitude loss through the adapters and cables
connected between the signal generator’s RF output and the mm-wave source module’s RF
input should be less than 1.5 dB.

Configure the Signal Generator

1. Turn on the signal generator’s line power. At power-up, the signal generator automatically does the

following:

senses the mm-wave source module

switches the signal generator’s leveling mode to external/source module

sets the mm-wave source module frequency and amplitude to the source module’s preset values

displays the RF output frequency and amplitude available at the mm-wave source module output

The

MMMOD

indicator in the

FREQUENCY

area and the

MM

indicator in the

AMPLITUDE

area of the signal

generator’s display indicate that the mm-wave source module is active

NOTE

For specific frequency/amplitude ranges, see the mm-wave source module specifications.

Advertising