background image

108

Chapter 5

Dual Arbitrary Waveform Generator
Using Waveform Markers

Waveform Marker Concepts

The Dual Arb mode of the signal generator has four markers that you can place on a waveform segment.
Marker 1 and Marker 2 provide auxiliary output signals to the rear-panel EVENT 1 and EVENT 2
connectors, respectively. Markers 3 and 4 are available only for custom-programmed waveforms, and they
provide auxiliary output signals to pins 19 and 18 of the rear-panel AUXILIARY I/O connector,
respectively. You can construct these output signals as a trigger signal to synchronize another instrument to a
given portion of a waveform. The following timing diagrams describe the effects of Markers 1 and 2 on the
state of the signal at the EVENT 1 and EVENT 2 rear panel connectors.

NOTE

If marker polarity selection is not be available in your version of the firmware, marker
polarity is always positive.

Figure 5-10

Table 5-1

Marker 1 and EVENT 1

Marker File
Bit 1

Waveform
point n

point n+1

point n+2

point n+3

. . .

Signal At EVENT
1 Connector

For Marker Polarity = Positive

For Marker Polarity = Negative

Marker File
Bit 1

Marker

Polarity

EVENT 1

Negative

Positive

±