To set the arb reference – Agilent Technologies E8247C PSG CW User Manual

Page 154

Advertising
background image

144

Chapter 6

Custom Arb Waveform Generator
Configuring Hardware

11. On the signal generator, press

Mode

>

Custom

>

Arb Waveform Generator

>

Digital Modulation Off On

until

On is highlighted.

This generates a waveform with the custom multicarrier state and the display changes to

Dig Mod

Setup: Multicarrier

.

During waveform generation, the

DIGMOD

and

I/Q

annunciators activate and the new custom

multicarrier state is stored in volatile ARB memory. The waveform should be modulating the RF carrier.

12. Press

RF On/Off

.

The externally single-triggered custom multicarrier waveform should be available at the signal
generator’s RF OUTPUT connector 100 ms after receiving a change in TTL state from low to high at the
PATT TRIG IN.

To Set the ARB Reference

Setting for an External or Internal Reference

1. Press

Custom

>

Arb Waveform Generator

>

More (1 of 2)

.

2. Press

ARB Reference Ext Int

to select either external or internal as the waveform sample clock reference.

If you select Ext, you must enter the reference frequency (250 kHz to 100 MHz) and apply the
reference signal to the rear-panel BASEBAND GEN REF IN.

If you select Int, the internal clock is used for the arbitrary waveform (ARB) frequency reference.

Setting the External Frequency

The external Arb reference frequency is only used when the

ARB Reference Ext Int

softkey has been set to Ext

(external).

1. Press

Custom

> Arb Waveform Generator >

More (1 of 2)

.

2. Press

Reference Freq

, enter a desired frequency (250 kHz to 100 MHz), and press

MHz

,

kHz

, or

Hz

.