RF Output Power Problems

RF Output Power Problems

Check the

RF ON/OFF

annunciator on the display. If it reads

RF OFF

, press

RF On/Off

to toggle the RF

output on.

RF Output Power too Low

1. Look for an

OFFS

or

REF

indicator in the

AMPLITUDE

area of the display.

OFFS

tells you that an amplitude offset has been set. An amplitude offset changes the value shown in the

AMPLITUDE

area of the display but does not affect the output power. The amplitude displayed is equal to

the current power output by the signal generator hardware plus the value for the offset.

To eliminate the offset, press the following keys:

Amplitude

>

More (1 of 2)

>

Ampl Offset

>

0

>

dB

.

REF

tells you that the amplitude reference mode is activated. When this mode is on, the displayed

amplitude value is not the output power level. It is the current power output by the signal generator
hardware minus the reference value set by the

Ampl Ref Set

softkey.

To exit the reference mode, follow these steps:

a. Press

Amplitude

>

More (1 of 2)

.

b. Press

Ampl Ref Off On

until Off is highlighted.

You can then reset the output power to the desired level.

2. If you are using the signal generator with an external mixer, see

“Signal Loss While Working with a

Mixer” on page 190

.

3. If you are using the signal generator with a spectrum analyzer, see

“Signal Loss While Working with a

Spectrum Analyzer” on page 192

.

The Power Supply has Shut Down

If the power supply is not working, it requires repair or replacement. There is no user-replaceable power
supply fuse. Refer to the Service Guide for instructions.