No modulation at the rf output – Agilent Technologies E8247C PSG CW User Manual

Page 203

Advertising
background image

Chapter 10

193

Troubleshooting

No Modulation at the RF Output

There are three power search modes: manual, automatic, and span.

When

Power Search

is set to Manual, pressing

Do Power Search

executes the power search calibration routine

for the current RF frequency and amplitude. In this mode, if there is a change in RF frequency or amplitude,
you will need to press

Do Power Search

again.

When

Power Search

is set to Auto, the calibration routine is executed whenever the frequency or amplitude of

the RF output is changed.

When

Power Search

is set to Span, pressing

Do Power Search

executes the power search calibration routine

over a selected range of frequencies at one time. The power search corrections are then stored and used
whenever the signal generator is tuned within the selected range of frequencies.

No Modulation at the RF Output

Check the

MOD ON/OFF

annunciator on the display. If it reads

MOD OFF

, press

Mod On/Off

to toggle the

modulation on.

Although you can set up and enable various modulations, the RF carrier is modulated only when you have
also set

Mod On/Off

to On.

For digital modulation, make sure that

I/Q Off On

is set to On.