Data Storage Problems

Data Storage Problems

Registers With Previously Stored Instrument States are Empty

The save/recall registers are backed-up by a battery when line power to the signal generator is not
connected. The battery may need to be replaced.

To verify that the battery has failed:

1. Turn off line power to the signal generator.

2. Unplug the signal generator from line power.

3. Plug in the signal generator.

4. Turn on the signal generator.

5. Observe the display for error messages.

If either error message

311 or

700 is stored in the error message queue, the signal generator’s battery

has failed.

6. Refer to the Service Guide for battery replacement instructions.

Saved Instrument State, but Register is Empty or Contains Wrong State

If you select a register number greater than 99, the signal generator automatically selects register 99 to save
the instrument state.

If the register number you intended to use is empty or contains the wrong instrument state, recall register 99:

Press

Recall

>

99

>

Enter

.

The lost instrument state may be saved there.