Signal generator locks up, Fail-safe recovery sequence – Agilent Technologies E8247C PSG CW User Manual

Page 208

Advertising
background image

198

Chapter 10

Troubleshooting
Signal Generator Locks Up

Signal Generator Locks Up

If the signal generator is locked up, check the following:

Make sure that the signal generator is not in remote mode (in remote mode, the

R

annunciator appears on

the display). To exit remote mode and unlock the front panel keypad, press

Local

.

Make sure that the signal generator is not in local lockout condition. Local lockout prevents front panel
operation. For more information on local lockout, refer to the Programming Guide.

Check for a progress bar on the signal generator display, which indicates that an operation is in progress.

Press

Preset

.

Cycle power on the signal generator.

Fail-Safe Recovery Sequence

Use the fail-safe recovery sequence only if the previous suggestions do not resolve the problem.

CAUTION

This process does reset the signal generator, but it also destroys the following types of data:

all user files (instrument state and data files)

DCFM/DC

Φ

M calibration data

persistent states

NOTE

Do not attempt to perform any other front panel or remote operations during the fail-safe
sequence.

To run the fail-safe sequence, follow these steps:

1. Hold down the

Preset

key while cycling power.

2. Continue to hold down the

Preset

key until the following message is displayed:

WARNING

You are entering the diagnostics menu which can cause unpredictable instrument
behavior. Are you sure you want to continue?

CAUTION

Carefully read the entire message! It may list additional risks with this procedure.