Configuring the rf output, Configuring a continuous wave rf output – Agilent Technologies E8247C PSG CW User Manual

Page 38

Advertising
background image

28

Chapter 2

Basic Operation
Configuring the RF Output

Configuring the RF Output

This section provides information on how to create continuous wave and swept RF (on

page 31

) outputs. It

also has information on using a mm-Wave source module to extend the signal generator’s frequency range
(see

page 47

).

Configuring a Continuous Wave RF Output

These procedures demonstrate how to set the following parameters:

RF output frequency

frequency reference and frequency offset (on

page 29

)

RF output amplitude (on

page 30

)

amplitude reference and amplitude offset (

page 30

)

Setting the RF Output Frequency

Set the RF output frequency to 700 MHz, and increment or decrement the output frequency in 1 MHz steps.

1. Return the signal generator to the factory-defined state: Press

Preset

.

NOTE

You can change the preset condition of the signal generator to a user-defined state. For these
examples, however, use the factory-defined preset state (set the

Preset Normal User

softkey

in the Utility menu to Normal).

2. Observe the

FREQUENCY

area of the display (in the upper left-hand corner).

The value displayed is the maximum specified frequency of the signal generator.

3. Press

RF On/Off

.

The

RF On/Off

hardkey must be pressed before the RF signal is available at the

RF OUTPUT connector. The display annunciator changes from

RF OFF

to

RF ON

. The maximum

specified frequency should be output at the RF OUTPUT connector (at the signal generator’s minimum
power level).

4. Press

Frequency

>

700

>

MHz

.

The 700 MHz RF frequency should be displayed in the

FREQUENCY

area of the display and also in the

active entry area.

5. Press

Frequency

>

Incr Set

>

1

>

MHz

.

This changes the frequency increment value to 1 MHz.