Chapter 3

61

Optimizing Performance

Using External Leveling

Configure the Signal Generator

1. Press

Preset

.

2. Press

Frequency

>

10

>

GHz

.

3. Press

Amplitude

>

0

>

dBm

.

4. Press

RF On/Off

.

5. Press

Leveling Mode

>

Ext Detector

.

This deactivates the internal ALC detector and switches the ALC input path to the front panel ALC
INPUT connector. The

EXT

indicator is activated in the

AMPLITUDE

area of the display.

NOTE

For signal generators with Option 1E1, notice that the

ATTN HOLD

(attenuator hold)

annunciator is displayed. During external leveling, the signal generator automatically
uncouples the attenuator from the ALC system for all external leveling points. While in this
mode, the RF output amplitude adjustment is limited to

20 to +25 dBm, the adjustment

range of the ALC circuitry. For more information, see

“External Leveling with Option 1E1

Signal Generators” on page 63

.

6. Observe the coupling factor printed on the directional coupler at the detector port. Typically, this value is

10 to

20 dB.

Enter the positive dB value of this coupling factor into the signal generator.

7. Press

More (1 of 2)

>

Ext Detector Coupling Factor

>

16

(or the positive representation of the value listed at

the detector port of the directional coupler) >

dB

.

Leveled output power is now available at the output of the directional coupler.

NOTE

While operating in external leveling mode, the signal generator’s displayed RF output
amplitude is affected by the coupling factor value, resulting in a calculated approximation of
the actual RF output amplitude. To determine the actual RF output amplitude at the point of
detection, measure the voltage at the external detector output and refer to

Figure 3-3

or

measure the power directly with a power meter.