Configuring AM

Configuring AM

In this example, you will learn how to generate an amplitude-modulated RF carrier.

To Set the Carrier Frequency

1. Press

Preset

.

2. Press

Frequency

>

1340

>

kHz

.

To Set the RF Output Amplitude

Press

Amplitude

>

0

>

dBm

.

To Set the AM Depth and Rate

1. Press the

AM

hardkey.

2. Press

AM Depth

>

90

>

%

.

3. Press

AM Rate

>

10

>

kHz

.

The signal generator is now configured to output a 0 dBm, amplitude-modulated carrier at 1340 kHz with
the AM depth set to 90% and the AM rate set to 10 kHz. The shape of the waveform is a sine wave. Notice
that sine is the default selection for the

AM Waveform

softkey, which can be viewed by pressing

(More 1 of 2)

.

To Turn on Amplitude Modulation

Follow these remaining steps to output the amplitude-modulated signal.

1. Press the

AM Off On

softkey to On.

2. Press the front panel

RF On Off

key.

The

AM

and

RF ON

annunciators are now displayed. This indicates that you have enabled amplitude

modulation and the signal is now being transmitted from the RF OUTPUT connector.