Test procedure: 6-4, Return loss con fi dence test 6-3 - 5, Return loss confidence test -3 – Anritsu ME7840A User Manual

Page 100: Setup -3

Advertising
background image

RETURN

RETURN

Step 5.

Press

Display

key and select:

DISPLAY MODE

SINGLE CHANNEL

RETURN

Step 6.

Press

Ch 4

key and then

Display

key.

Step 7.

Select:

GRAPH TYPE

LOG MAGNITUDE

Step 8.

Press

Ch

1 key and select

LOG MAGNITUDE.

Test Procedure:

Insert the Calibration Component Coefficients disk into the MS462XC
floppy disk drive.

Step 1.

Press Cal key and select:

COMPONENT UTILTIES

INSTALL KIT INFO FROM FLOPPY DISK

RETURN

Step 2.

Follow the prompts and select:

PERFORM CAL: 2 PORT

CAL METHOD: STANDARD

LINE TYPE: COAXIAL

SELECT CALIBRATION TYPE: REFLECTION ONLY –
PORT 1 ONLY

FREQUENCY RANGE OF CALIBRATION: 800 MHz TO
2400 MHz

SELECT CALIBRATION DATA POINTS: NORMAL

DATA POINTS: 401 POINTS

PORT 1 CONNECTOR TYPE: N (F)

LOAD TYPE: BROADBAND LOAD

Step 3.

Select

START CAL

soft key to begin calibration.

Step 4.

Install calibration device per instruction on the display. Select
the appropriate soft key to measure the calibration device.

Step 5.

When the “CALIBRATION SEQUENCE COMPLETED” mes-
sage is displayed, press the

Enter

key to continue.

RETURN LOSS CONFIDENCE TEST

PERFORMANCE VERIFICATION

6-4

ME 7840A OMM