Technical details cal specification files – Anritsu ME7840A User Manual

Page 114

Advertising
background image

A-4

ME 7840A OMM

TECHNICAL DETAILS

CAL SPECIFICATION FILES

II CALIBRATION SPECIFICATIONS, TYPES

1.

pow (Power Level)

Equipment Required: Power meter connected to Dedicated GPIB port

Power sensor with appropriate frequency range
Throughline

Header Section Keywords

ncal

Number of calibration sequences of this type to run. For type = pow, each
sequence consists of:

linear cal (for power sweeps)
flat power cal
receiver cal

NOTE: Since this attempts to be a "global" power level calibration, ncal should

get a value of 1.

npts

Number of frequency points to use during the calibration sequence. Increasing
this number will increase calibration time.
Legal values are (3, 15, 51, 101, 201, 401, 801, 1601 only)
NOTE: Illegal value causes software to default to 51.

if

IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000
NOTE: Illegal value causes software to default to 1000.

avgs

Averages: (Optional). Software will set averaging to value (and turn on) during cal.

pow

Source setting for each source during flat power cal.

target

Target power setting for each source during flat power cal.

noffset

Nominal offset value for use during linear cal.
NOTE: During linear cal the software attempts to calibrate the sources from

-15 dBm to +5 dBm at the input reference plane.

switch

Setting for test set. Correponds to 0

-

15 settings.

Table A-1. Technical Details of Calibration Specification Files (3 of 9).

Header

POW_START

POW_END

Details