Technical details cal specification files – Anritsu ME7840A User Manual

Page 116

Advertising
background image

TECHNICAL DETAILS

CAL SPECIFICATION FILES

A-6

ME 7840A OMM

2.

sp and Hot S

22

(S Parameters & Hot S

22

)

Header Section Keywords:

ncal

Number of calibration sequences of this type to run.

npts

Number of frequency points to use during the calibration sequence. Increasing
this number will increase calibration time.
Legal values are (3, 15, 51, 101, 201, 401, 801, 1601 only)
NOTE: Illegal value causes software to default to 51.

con

Connector sypes for Port 1 and Port 2: Options:

MN, FN - Male/Female N type
MK, FK - Male/Female K
MS, FS - Male/Female SMA

if

IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000
NOTE: Illegal value causes software to default to 1000.

avgs

Averages: (Optional). Software will set averaging to value (and turn on) during cal.

pow

Source power setting: during cal, Source 1.

switch

Setting for test set. Correponds to 0

-

15 settings.

Detail Section Keywords:

CAL, 1

First calibration

cal type

Type of calibration to perform:

12 - 12 term
RR - Reflection Reverse
RF - Reflection Forward
8T - 1 Path 2 Port Forward
8R - 1 Path 2 Port Reverse
FT - Transmission Frequency Response

<name>

Name for Hot S

22

<hot>

For Hot S

22

only. Value = offset in kHz of 2nd Source

Table A-1. Technical Details of Calibration Specification Files (5 of 9).

Header

SP_START

SP_END

Details

Advertising