Hot s22 4-8, Hot s22 operations, calibration – Anritsu ME7840A User Manual

Page 50

Advertising
background image

4-7

HOT S22

The “B_HotS22.txt” file provides prompts for setting up a Hot S

22

specifi-

cation file. Hot S

22

is a return loss measurement of an amplifier’s output

port (Port 2) while stimulus is applied to its input port. The prompts
cause the Scorpion to be placed into a state where it is ready to make cali-
brated measurements on the user’s AUT. At the completion of the pro-
gram, the state information is saved to a *.cal file. To run this program,
proceed as follows.

Step 1.

Select “Tools” then “Run Calibration Procedure.”

Step 2.

Double-click on “C_Hot S22.txt” to begin the Hot S

22

cal file

setup. Follow the prompt and connect open, short, and load to
Port 2 as directed.

HOT S22

OPERATIONS, CALIBRATION

4-8

ME 7840A OMM