S-parameter test: k factor 5-7, Operations, measurement s-parameter test: k factor – Anritsu ME7840A User Manual

Page 68

Advertising
background image

5-6

S-PARAMETER
TEST: K FACTOR

This test describes the K Factors measurement. This measurement is
described in Chapter 1, paragraph 1-10.

Procedure.

Prepare the ME7840A as shown in Chapter 3.

Enter the appropriate test values in the displayed
User Input screen. The input fields are described
below: The inputs required for this test are de-
scribed below.

n

S-Parameter: Drop down options. K Factor for

this measurement.

n

Data Points: Drop down options

- number of data

points in frequency sweep.

n

Frequency 1 (MHz): Lower frequency or start fre-

quency.

n

Frequency 2 (MHz): Upper frequency or stop fre-

quency.

n

Input Power (dBm):Input power level, in dBm.

n

IF Bandwidth (Hz):Intermediate frequency band-

width for test, in Hertz.

n

Test Set Setting: Byte value for setting test set

attenuation and source switch (below).

OPERATIONS, MEASUREMENT

S-PARAMETER TEST: K FACTOR

ME7840A OMM

5-7

Internal Source

Value

External Source

Value

0 dB

8

0 dB

0

10 dB

9

10 dB

1

20 dB

10

20 dB

2

30 dB

11

30 dB

3

40 dB

12

40 dB

4

50 dB

13

50 dB

5

60 dB

14

60 dB

6

70 dB

15

70 dB

7