Power level operations, calibration – Anritsu ME7840A User Manual

Page 48

Advertising
background image

NOTE

The linear power calibration takes up-to-5 minutes to
complete. It calibrates the output power at the refer-
ence plane for both internal sources (power out of
Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci-
fication file. When the test completes, the “Next” but-
ton will become available.

Step 5.

Press “Next” to calibrate flat power response (below).

NOTE

The flat power calibration also takes less than 1 min-
ute to complete. It calibrates the output power at the
reference plane for both internal sources (power out
of Port 1 then Port 3) according to the target power
across the frequency range set in the calibration speci-
fication file. When the test completes, the “Next” but-
ton will become available.

POWER LEVEL

OPERATIONS, CALIBRATION

4-6

ME 7840A OMM