Step 4.

Connect power sensor to MS462XC

Port 1

and select

START

VERIFICATION

softkey.

Step 5.

Verify that Source 1 passes this test.

Step 6.

Press the

SELECT SOURCE

soft key to select 2

Step 7.

Connect power sensor to MS462XC

Port 3

and select

START

VERIFICATION

soft key.

Step 8.

Verify that Source 2 passes this test.

Step 9.

Select the menu soft key as follows to exit the TROUBLE-
SHOOTING mode:

RETURN

RETURN

FINISHED, RECOVER FROM TROUBLESHOOTINGT

Step 10.

Reconnect cables between MS462XC

Port 1

and

Port

1 of

MS4782 Test Set and between

Port 3

of MS462XC and

Port 3

of MS4782 Test Set.

6-5

RETURN LOSS
CONFIDENCE
TEST

This test verifies the accuracy of return loss measurements.

Setup

Turn on the MS462XC VNMS and MS4782X Test Set and allow them to
warm up for 30 minutes.

Step 1.

Press

Default

key then

0

key.

Step 2.

Press

Seq

key and .select:

TTL I/O.

PARALLEL OUT SETUP

SET PARALLEL OUT PORT (0-255)

Step 3.

Change the value from 0 to 8.

Step 4.

Select:

PERFORMANCE VERIFICATION

RETURN LOSS CONFIDENCE TEST

ME7840A OMM

6-3

CAUTION

The TROUBLESHOOTING function must be properly exited to restore
normal

NOTE

The worst case test result
will be displayed on the
screen when the instru-
ment fails this test only.
The failed test result is also
recorded in theService
Log.

Advertising