Receiver display linear ity 6-8 - 10, Receiver display linearity -8, Setup: -8 – Anritsu ME7840A User Manual

Page 105

Advertising
background image

Step 9.

Toggle

C.W. MODE OFF

soft key to

C.W. MODE ON

.

Step 10.

Change C.W. Frequency to 1 GHz.

Step 11.

Press

Config

key and select:

DATA POINTS

POINTS DRAW IN C.W. – 51 POINT(S)

Step 12.

Press

Avg

key and select:

SELECT I.F. BANDWIDTH

I. F. BW 10 Hz

Step 13.

Press

Power

key and select SOURCE 1 SETUP.

Step 14.

Adjust

SOURCE 1 POWER

so that the power meter readout is

0 dBm ± 0.1 dB.

Step 15.

Press

Ch 3

key.

Step 16.

Press

Meas

key and select

USER DEFINED

.

Step 17.

Change

RATIO

to

b

2/

1.

Step 18.

Select

S21/USER 1

soft key to use user defined parameter.

Step 19.

Press Display key and select:

GRAPH TYPE

LOG MAGNITUDE

RETURN

TRACE MEMORY

Step 20.

Allow the trace to sweep twice. Select:

STORE DATA TO MEMORY

VIEW DATE (/) MEMORY

Step 21.

Press

Marker

key and select

MARKER READOUT.

Step 22.

Use the soft key to turn on Marker 1. Then use the numeric
data entry key to change

POINT

to

25

.

Step 23.

Press

Marker

key again.

Step 24.

Use the soft key to

SCREEN DISPLAY ON

.

PERFORMANCE VERIFICATION

RECEIVER DISPLAY LINEARITY

ME7840A OMM

6-9

NOTE

Ensure that the Power
Sensor Cal Factor is set for
1 G H z .

R e f e r t o t h e

ML24XXA Operational

Advertising