System dynamic range 6-6, Test procedure: 6-6, System dynamic range performance verification – Anritsu ME7840A User Manual

Page 102

Advertising
background image

6-6

SYSTEM DYNAMIC
RANGE

This test verifies the PATS system dynamic range.

Test Procedure:

Step 1.

Press

Default

key then

0

key of MS462XC to reset the system.

Step 2.

Press

Seq

key.

Step 3.

Select:

TTL I/O.

PARALLEL OUT SETUP

SET PARALLEL OUT PORT (0-255)

Step 4.

Change the value from 0 to 8.

Step 5.

Select:

RETURN

RETURN

Step 6.

Press

Ch 3

key.

Step 7.

Press

Display

key and select:

DISPLAY MODE

SINGLE CHANNEL

RETURN

GRAPH TYPE

LOG MAGNITUDE

RETURN

Step 8.

Press

Cal

key.

Step 9.

Follow the prompts and select::

SYSTEM DYNAMIC RANGE

PERFORMANCE VERIFICATION

6-6

ME 7840A OMM

Offset Termination

Measured S

11

Value

Measured S

22

Value*

Specification

SC5237

6

±

0.4 dB

SC5270

20

±

1.0 dB

* Apply to Test Set with Hot S

22

Capability only.

Table 6-1. Return Loss Specifications

Advertising