CAL SPECIFICATION FILES

TECHNICAL DETAILS

ME7840A OMM

A-9

4.

harm (Harmonics):

Header Section Keywords:

ncal

Number of calibration sequences of this type to run.

if

IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000

avgs

Averages: (Optional). Software will set averaging to value (and turn on) during cal.

pow

Start power level for sweep: Set on for Source 1.

switch

Setting for test set. Correponds to 0

-

15 settings.

Detail Section Keywords:

CAL, 1

First calibration

name

Name of *.cal file to save when calibration sequence completes.

f1

Main tone, MHz

f2

Maximum display range, MHz
NOTE: Minumum display range is automatically set to 100 MHz.

Example

Refer to Figure A-2 (page A-12).

Table A-1. Technical Details of Calibration Specification Files (8 of 9).