Harmonics 5-29, Operations, measurement harmonics – Anritsu ME7840A User Manual

Page 90

Advertising
background image

5-10

HARMONICS

The harmonics test measures and displays the harmonics in the device
under test.

Procedure

Prepare the ME7840A as shown in Chapter 3.

Enter the appropriate test values in the displayed
User Input screen. The input fields are described
below:

n

Measurement Type: Already set to “HA” for Har-

monics – input disabled.

n

Start Frequency (MHz): Lower frequency to be

displayed. Typically set to a value less than or
equal to the main tone.

n

Stop Frequency (MHz): Upper Frequency to be

displayed. Set to a value less than the system’s
maximum frequency (for example, for MS4623C
set to a max of 5000 MHz).

n

Main Tone (MHz): CW Frequency setting for

source 1.

n

Input Power: Input RF power to AUT.

n

Test Set Setting: Byte value for setting test set

attenuation and source switch:

OPERATIONS, MEASUREMENT

HARMONICS

ME7840A OMM

5-29

Internal Source

Value

External Source

Value

0 dB

8

0 dB

0

10 dB

9

10 dB

1

20 dB

10

20 dB

2

30 dB

11

30 dB

3

40 dB

12

40 dB

4

50 dB

13

50 dB

5

60 dB

14

60 dB

6

70 dB

15

70 dB

7