46

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

2

Hardware Setup and Configuration

Measurement Procedure

5

Tune the Spectrum Analyzer to the following:

Frequency: 7.5 MHz

Amplitude: 10 dBm

Span: 100 kHz

6

Tune the downconverter to the following:

Level: Attenuator: 20 dBm

RF Enabled: Enabled

Input RF: Signal Source: 50 MHz Cal

Frequency: 50 MHz

7

Verify that a 7.5 MHz signal is present on the spectrum analyzer.