Proper connector torque, Connector wear and damage, Sma connector precautions – Agilent Technologies N8201A User Manual

Page 97

Advertising
background image

Preventive Maintenance

5

Agilent N8201A Performance Downconverter Synthetic Instrument Module, 3 Hz to 26.5 GHz

93

Do not bend cables near the connectors.

If any of the cables will be flexed repeatedly, buy a back-up cable. This will allow
immediate replacement and will minimize system down time.

Before connecting the cables to any device:

Check all connectors for wear or dirt.

When making the connection, torque the connector to the proper value.

Proper Connector Torque

Provides more accurate measurements

Keeps moisture out of the connectors

Eliminates radio frequency interference (RFI) from affecting your measurements

The torque required depends on the type of connector. Refer to

Table 4

.

Never exceed the recommended torque when attaching cables.

Connector Wear and Damage

Look for metal particles from the connector threads and other signs of wear (such as
discoloration or roughness). Visible wear can affect measurement accuracy and
repeatability. Discard or repair any device with a damaged connector. A bad connector can
ruin a good connector on the first mating. A magnifying glass or jeweler’s loupe is useful
during inspection.

SMA Connector Precautions

Use caution when mating SMA connectors to any precision 3.5 mm RF connector. SMA
connectors are not precision devices and are often out of mechanical tolerances, even
when new. An out-of-tolerance SMA connector can ruin a 3.5 mm connector on the first
mating
. If in doubt, gauge the SMA connector before connecting it. The SMA center
conductor must never extend beyond the mating plane.

Table 4

Proper Connector Torque

Connector

Torque cm-kg

Torque N-cm

Torque in-lbs

Wrench P/N

Type-N

52

508

45

hand tighten

3.5 mm

9.2

90

8

8720-1765

SMA

5.7

56

5

8710-1582

Advertising