Level

Sets the input attenuation. The input attenuation can be set from 0 to 70 dB in 2 dB
steps.The input impedance is set to 50 ohms. Input attenuation is used to minimize
compression caused by a signal level that is too high in amplitude.

Input RF

Accesses the functions to select the input signal source (choices are RF, 50 MHz Cal, or
External Mixer), enable a preamplifier, and to set the coupling to either AC or DC.

IF Output

Accesses the functions to set the IF Bandwidth and IF Frequency.

Reference Oscillator

Allows you to use either the internal or an external reference oscillator for making
measurements.

Reference Output

Switches the 10 MHz out signal on the front panel of the N8201A performance
downconverter On and Off.

Reference Input Source

Specifies whether to use an External or Internal source for the reference oscillator. If set to
External, a value must be specified for External Frequency with limits from 1 MHz to
30 MHz.

Table 3

Grouped Values

IF Output

IF Bandwidth

Limits

IF Frequency

Limits

Input Impedance

Limits

IF Output 1

0 to 3 MHz

21.4 MHz

50 Ohm

IF Output 2

100 MHz

321.4 MHz

50 Ohm

IF Output 3

10 MHz

7.5 MHz

50 Ohm