1 operation modes, Smode setting the serial interface, Open & close – Vaisala HMP230 User Manual

Page 101

Advertising
background image

HMP230 SERIES

M210225en-B

Appendix 1: Serial commands

95

3.1

Operation modes

SMODE Setting the serial interface

SMODE xxxx<cr>

xxxx

=

STOP, RUN or POLL

In STOP mode: measurements output only by command, all commands can be

used

In RUN mode: outputting automatically, only command S can be used

In POLL mode: measurements output only with command SEND. When in

POLL mode, the output state is changed as follows:

OPEN aa <cr>

SMODE xxxx<cr>

aa

=

address of the transmitter

xxxx

=

STOP, RUN or POLL

The OPEN command sets the bus temporarily in STOP MODE so that the
SMODE command can be given. For example:

>SMODE <cr>

which mode is in use at the moment

Serial mode

: STOP

>SMODE STOP <cr>

setting STOP mode

Serial mode

: STOP

OPEN & CLOSE

OPEN nn <cr>

nn

=

address of the transmitter (0...99)

CLOSE <cr>

In STOP mode: command OPEN has no effect, CLOSE sets the transmitter in

POLL mode

In POLL mode: command OPEN sets the transmitter temporarily in STOP

mode, command CLOSE returns the instrument to POLL
mode