the connection returns the transmitter to POLL state. In POLL state the
transmitter outputs measurement only when requested (command SEND aa). If
the user has forgotten the address of the transmitter and the transmitter does
not have a display, the transmitter has to be reverted to the factory settings
(see Chapter 4.4.1). If the transmitter has a display, the settings can be
checked through it.

6.3.6.1. Setting the serial interface operation mode

Select Mode in the main menu; the following is displayed:

Select Serial output:

The currently valid setting flashes. Select the desired mode with the
arrow keys and press ENT. After this the program returns to the Mode
Menu.

When RUN mode is selected, the currently valid output interval is
displayed:

The output interval setting can be changed as follows:

press CL

the number starts flashing

if the interval needs to be changed, press CL again and enter the new
interval; otherwise press ENT

the unit (s, min, h) starts flashing

the unit can be changed with the arrow keys and acknowledged with
ENT

after this the program returns to Mode menu