HMP230 SERIES

M210225en-B

User's Guide

77

9.8.

Electromagnetic compatibility

HMP230 series transmitters fulfill the standard EN 61326-1:1997 +Am
1:1998, Electrical equipment for measurement, control and laboratory use -
EMC requirements; Industrial environment (Generic environment for HMP233
with low temperature sensor cable).

Test methods

Emission
Radiated emissions

CISPR16 class B (CISPR22 Class B)

Immunity
Electrostatic discharge (ESD)

EN/IEC 61000-4-2

EM field

EN/IEC 61000-4-3
Industrial environment: 10 V/m
Generic environment: 3 V/m

EFT Burst

EN/IEC 61000-4-4

Surge

EN/IEC 61000-4-5

Conducted RF

EN/IEC 61000-4-6

NOTE

These tests are valid with the following specified sensor
protections only.

Type:

Sensor protection/notes:

HMP231

sintered filter (part no. 16452), PPS grid &
stainless steel netting (part no. 16720)

HMP233
high temperature (120

°

C)

PPS grid & stainless steel netting
(part no. 16720)

low temperature (80

°

C)

sintered filter (part no. 16452)
PPS grid & stainless steel netting
(part no. 16720)

HMP234

sintered filter (part no. 16452)

HMP235

sintered filter (part no. 16452), PPS grid &
stainless steel netting (part no. 16720)

HMP237

sintered filter (part no. 16452), PPS grid &
stainless steel netting (part no. 16720)

HMP238

sintered filter (part no. 16452)

Voltage proof, AC: DC supply (+ or -) to housing 250 VAC, 1 minute (300 k

and 15 nF parallel)