HMP230 SERIES
User's Guide

M210225en-B

62

7.2.2. Two point temperature calibration

7.2.2.1. With serial commands

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Disconnect the security lock jumper!

Check the transmitter against the lower reference.

Give command CT <cr>, enter the first point value and press <cr>:

>CT <cr>
T : xx.x Ref1 ? yy.y <cr>
Press any key when ready

If you want to see how the sensor stabilizes to the reference
temperature, enter c <cr> instead of the first reference:

T : 0.90 Ref1 ? c <cr>
T : 0.55 Ref1 ? c <cr>
T : 0.55 Ref1 ? 0.0 <cr>
Press any key when ready...

Change the temperature and check the transmitter against the higher
reference.

Check that the reading corresponds with the reading of the reference
instrument. If not, adjust the second point

Press any key, enter the second point value and press <cr>.

T : xx.x Ref2 ? yy.y <cr>

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

7.2.2.2. With display/keypad commands

Leave the reference instrument and the transmitter for at least 4 hours
in the same space so that their temperatures have time to equalize.
Remove the filter cap prior to calibration.

Disconnect the security lock jumper!

Check the transmitter against the lower reference.

Select Cali in the main menu and then T cal; select two-point
calibration T 2 point cal. Change the first point reading with the
arrow keys and press ENT.

Change the temperature and check the transmitter against the higher
reference.