Vaisala HMP230 User Manual

Page 71

Advertising
background image

HMP230 SERIES

M210225en-B

User's Guide

65

the LED on the left flashes; set the low end of channel 1 with the ar-
row keys and press ENT

the second LED from the left flashes; set the high end of channel 1
with the arrow keys and press ENT

the LED on the left flashes; set the low end of channel 2 with the ar-
row keys and press ENT

the second LED from the left flashes; set the high end of channel 2
with the arrow keys and press ENT

The analogue outputs are calibrated to ensure that outputs are correctly scaled:
for example, when the output is scaled to 4...20 mA, the low end of the scale
is 4 mA and high end 20 mA exactly. However, when 0... 20 mA output is
used, the output cannot be adjusted to exactly 0 mA, but to 50 µA. When 0...1
V, 0...5 V or 0...10 V output is in use, the output is adjusted to 50 mV. The
following table summarizes the correct output values.

Summary of the correct output values:

Output scale:
0...20 mA

4...20 mA

0...1 V

0...5 V

0...10 V

low end:

50 µA

4 mA

50 mV

50 mV

50 mV

high end:

20 mA

20 mA

1 V

5 V

10 V

Advertising