Technical data – Vaisala HMP230 User Manual

Page 77

Advertising
background image

HMP230 SERIES

M210225en-B

User's Guide

71

9.

TECHNICAL DATA

9.1.

Relative humidity

Measuring range

0...100 %RH

Accuracy (including nonlinearity and repeatability)

maximum achievable accuracy when calibrated against high quality,
certified humidity standards:

±1 %RH (0...90 %RH)
±2 %RH (90...100 %RH)

when calibrated against salt solutions (ASTM E104-85):

±2 %RH (0...90 %RH)
±3 %RH (90...100 %RH)

Response time (90 %) at +20 °C in

15 s

still air (with sintered filter)

Sensors

HUMICAP

®

K:

humidity sensor with blue dots
on the sensor

HUMICAP

®

KC:

combined RH and T sensor
(part no.18258HM)

9.2.

Temperature

Measuring range

HMP231

-40...+60 °C

HMP233

-40...+80 / +120 °C

HMP234/235/237/238

-40...+180 °C

Typical accuracy of electronics at
+20 °C (+68 °F)

±0.1 °C

Typical temperature dependence of
electronics

0.005 °C/°C

Sensor

Pt 100 RTD 1/3 Class B IEC 751

9.3.

Calculated variables

Typical ranges - HMP231

dewpoint temperature

-40...+60 °C

mixing ratio

0...160 g/kg dry air

absolute humidity

0...160 g/m

3

wet bulb temperature

0...+60 °C

enthalpy

-40…+460 kJ/kg