HMP230 SERIES

M210225en-B

User's Guide

69

8.5.

Measurement of output currents using test points

If a current output has been connected e.g. to a process computer, the output
current cannot be measured at the output connector X2 without disconnecting
the external load. The output current can, however, be measured at test points
CH1+/CH1- and CH2+/CH2- without disconnecting the output wires. These
test points can therefore be used in one point offset correction against an accu-
rate reference or in checking the current output without disconnecting the
analogue output from the process.

X2

OPENED COVER OF THE HMP230

X15

CH2

CH1 +

+

C

H

1

+

C

H

1

-

C

H

2

+

C

H

2

-

Figure 32 Location of the CH1 and CH2 test points

CH1+

CH1-

TEST POINTS

CH1+

CH1-

mA

R

L

X2

Figure 33 Circuit diagram of the analogue output current test points