Accuracy of wet bulb temperature (°c) – Vaisala HMP230 User Manual

Page 79

Advertising
background image

HMP230 SERIES

M210225en-B

User's Guide

73

Accuracy of wet bulb temperature (°C)

Relative humidity

Temp.

10

20

30

40

50

60

70

80

90

100

-40

0.20

0.20

0.20

0.20

0.20

0.20

0.20

0.20

-

-

-20

0.21

0.21

0.22

0.22

0.22

0.22

0.23

0.23

-

-

0

0.27

0.28

0.28

0.29

0.29

0.29

0.30

0.30

0.31

0.31

20

0.45

0.45

0.45

0.44

0.44

0.44

0.43

0.43

0.42

0.42

40

0.84

0.77

0.72

0.67

0.64

0.61

0.58

0.56

0.54

0.52

60

1.45

1.20

1.03

0.91

0.83

0.76

0.71

0.67

0.63

0.60

80

2.23

1.64

1.32

1.13

0.99

0.89

0.82

0.76

0.72

0.68

100

3.06

2.04

1.58

1.31

1.14

1.01

0.92

0.85

0.80

0.75

120

3.85

2.40

1.81

1.48

1.28

1.13

1.03

0.95

0.88

0.83

140

4.57

2.73

2.03

1.65

1.41

1.25

1.13

1.04

0.97

0.91

160

5.25

3.06

2.25

1.82

1.55

1.37

1.24

1.13

1.05

0.99

Accuracy of enthalpy (kJ/kg) at ambient pressure 1013 mbar

Relative humidity

Temp.

10

20

30

40

50

60

70

80

90

100

-40

0.21

0.21

0.21

0.21

0.21

0.21

0.21

0.21

-

-

-20

0.24

0.25

0.25

0.25

0.26

0.26

0.26

0.27

-

-

0

0.4

0.42

0.43

0.45

0.46

0.47

0.49

0.5

0.52

0.53

20

0.98

1.03

1.08

1.13

1.18

1.23

1.29

1.34

1.39

1.45

40

2.7

2.87

3.04

3.22

3.4

3.59

3.78

3.98

4.18

4.39

60

7.2

7.82

8.48

9.19

9.96

10.79

11.69

12.65

13.7

14.84

80

18.04

20.72

23.88

27.66

32.22

37.77

44.61

53.17

64.02

78.06

100

43.96

57.64

77.98

109.93

164.1

266.37

494.62

1180.46

-

-

120

111.6

203.93

470.52

-

-

-

-

-

-

-

The formulas used in calculation of dewpoint, mixing ratio, absolute humidity
and enthalpy are presented in Appendix 6.

9.4.

Pressure

Pressure range:

HMP234

0...10 MPa (0...100 bar)

HMP237

0...1MPa (0...10 bar)

HMP238

0…4 MPa (0…40 bar)

9.5.

Analogue outputs

Two analogue outputs selectable

0...20 mA

4...20 mA

0...1 V

0...5 V

0...10 V

Typical accuracy of analogue output
at +20 °C

±0.05 % full scale

Typical temperature dependence of
analogue output

0.005 %/°C full scale

Advertising