Vaisala HMP230 User Manual

Page 46

Advertising
background image

HMP230 SERIES
User's Guide

M210225en-B

40

Selecting stop bits:

Full duplex/half duplex:

The processor does not allow the following combinations:

no parity, 7 data bits, 1 stop bit: if this combination is given the
HMP230 program will change the number of stop bits to 2

even or odd parity, 8 data bits, 2 stop bits: if this combination is given
the program changes the number of stop bits to 1

NOTE

The serial bus settings become effective only after reset.

6.3.5.3. Setting the transmitter address

Address is used when more than one transmitter is connected to one serial bus;
this way, it is possible to communicate with one transmitter at a time.

Select Addr in the main menu; the following is displayed:

Pressing ENT returns the program to the main menu.

Pressing CL deletes the old address; enter the new address with the
arrow keys.

Advertising