HMP230 SERIES
Appendix 1: Serial commands

M210225en-B

96

When more than one transmitter is connected to the same serial bus, the POLL
mode makes it possible to communicate with the transmitters. For example, a
relative humidity calibration is performed at transmitter 2 (<bel> = ASCII 7):

>OPEN 2 <cr>
<cr><lf>

'HMP

nn

line

opened

for

operator

commands'

<cr><lf><lf><bel>
>CRH <cr>
...
>CLOSE <cr>
<cr><lf> 'line closed' <cr><lf>