HMP230 SERIES
User's Guide

M210225en-B

56

7.1.2.1. With serial commands

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Disconnect the security lock jumper!

Insert the sensor head into a measurement hole of the LiCl chamber in
the humidity calibrator.

Wait at least 30 minutes.

Give command CRH <cr>, enter the first point value and press <cr>.

>CRH <cr>
RH : xx.x Ref1 ? yy.y <cr>
Press any key when ready...

If you want to see how the sensor stabilizes to the humidity in the
calibrator, enter c <cr> instead of the first reference:

RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

Insert the sensor head into a measurement hole of the NaCl chamber.

Wait at least 30 minutes.

Press any key and enter the second point value and press <cr>.

RH : xx.x Ref2 ? yy.y <cr>

The stabilization of the sensor can be monitored by entering c <cr>
instead of the reference value.

7.1.2.2. With display/keypad commands

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

Disconnect the security lock jumper!

Insert the sensor head into the measurement hole of the LiCl chamber.

Wait at least 30 minutes.

Select Cali in the main menu and then RH cal; select Not changed
and then two-point calibration RH 2 point cal. Change the first point
reading with the arrow keys to correspond the reference humidity and
press ENT; pressing an arrow once changes the reading by 0.05 %RH.