Vaisala HMP230 User Manual

Page 27

Advertising
background image

HMP230 SERIES

M210225en-B

User's Guide

21

slip the brass disk that has the bigger hole in it over the braid so that
it rests against the cable insulation

slip the other brass disk over the wires to the middle of the braid

(If the cable diameter is less

than 7mm, use a shrinking

tube or an adhesive tape)

cable

braid

flexible wires 0.5 mm²

(AWG 20), stranded wires

recommended

D = Ø 7...10 mm

3

brass
disks

25

rubber
ring

nut

brass disks

140

165

shielding tube

braid

push back the braid and press it between the two brass disks to
achieve a full 360° grounding; the fold between the disks should have
the same diameter as the brass disks

secure the braid with a shielding tube

insert the wires into the transmitter housing through the gland.

tighten the nut

connect the wires into the screw terminals and fasten a cable tie
around the wires

transmitter housing

cable tie

gland

brass disks

rubber ring

nut

Advertising