Vaisala HMP230 User Manual

Page 23

Advertising
background image

HMP230 SERIES

M210225en-B

User's Guide

17

See Figure 16 through Figure 18 for detailed description of installation
through the ball valve assembly. This installation is possible provided that
the process pressure is less than 10 bars.
Note also that if the sensor head is
installed in a process pipe, the nominal size of the pipe must be at least 1 inch.

STEP I: mount the probe with the ball valve assembly closed; tighten the

clasp nut manually

fitting body
R1/2 cone, sealed

bushing
R1/2 cone
sealed

bushing R1/2 cone/G1/2(40 bar)
e.g. Camozzi 2520-1/2-1/2

ball valve 1/2" (40 bar)
e.g. Atlas Copco:BAL-1A 15 (G1/2)

nipple
R1/2 cone
sealed

(the bushing serves for
moving the probe (sinter)
to such a distance from the
ball valve that the valve
can be closed)

>30 mm

clasp nut

Figure 16 Installing the probe through the ball valve assembly; step 1.