HMP230 SERIES
User's Guide

M210225en-B

64

>ACAL <cr>
Ch1

U1 ( V ) ? 2.233 <cr>

Ch1

U2 ( V ) ? 8.25 <cr>

Ch2

U1 ( V ) ? 2.218 <cr>

Ch2

U2 ( V ) ? 8.23 <cr>

7.3.2. With display/keypad commands

Disconnect the security lock jumper!

Connect an ammeter/voltmeter to the output of channel 1, select Cali
in the main menu and Analog outputs in the Cali menu. The
following is displayed (the quantity can be either mA or V):

Enter the measured lower end current/voltage on channel 1.

Enter the measured upper end current/voltage on channel 1.

Connect the meter to the output of channel 2 and enter the measured
lower end current/voltage on channel 2.

Enter the measured upper end current/voltage on channel 2.

7.3.3. With LED commands

If both the analogue outputs and humidity/temperature channels are calibrated,
the analogue outputs should be calibrated first. This applies only when the
calibrations are made using the LED commands!

connect an ammeter/voltmeter to the analogue outputs (connector X2)

Disconnect the security lock jumper!

Give command ¡¡ll.