Anritsu MG3641A User Manual

Page 103

Advertising
background image

SECTION 5 MEASUREMENT

STEP

PROCEDURE

7.

Turn on the output of MG3641A/MG3642A

2, place MG3641A/MG3642A 2 into the relative level mode,

and set the output level resolution to 1 dB.

8.

Set the frequency step size of MG3641A/MG3642A

2 to 20 kHz.

9.

Turn the rotary knob to adjust the output level so that the noise output of the receiver is set to V

N

dB

obtained in step 2 each time the Edit [^] key of MG3641A/MG3642A

2 is pressed. The value indicated by

the output level indicator is the disturbing input level separated from the desired wave by +

∆f × n.

10.

Return the frequency and output level of MG3641A/MG3642A

2 in the state set in step 3 (frequency and

output level indicators set to +0).

11.

Turn the rotary knob to adjust the output level so that the noise output of the receiver is set to V

N

dB

obtained in step 2 each time the Edit [

] key of MG3641A/MG3642A 2 is pressed. The value indicated by

the output level indicator is the disturbing input level separated from the desired wave by –

∆f × n.

12.

The following sensitivity suppression characteristics are obtained in steps 9 and 11:

>

80

60

40

20

0

20

40

60

80

120

100

80

60

40

20

Disturbing wave input level (dB)

Disturbing frequency offset [kHz]

Advertising