3 performance test, Caution – Anritsu MG3641A User Manual

Page 171

Advertising
background image

SECTION 7 PERFORMANCE TEST

7-3

7.3

Performance Test

To confirm performance after the MG3641A/MG3642A acceptance test, periodical inspection, or repair, check the

following items for performance test:

• Output frequency

• Output level frequency characteristics

• Output level accuracy

• FM deviation, FM distortion

• AM modulation factor, AM distortion

The performance test is executed according to the ordinary use state; therefore, the user need not adjust the inside of

the device like the calibration. For important items, the performance test should be performed periodically for preventive

maintenance.
Use the performance test result sheet in Appendix E to record the test results.

CAUTION

For performance test, the tested unit and measuring instruments must sufficiently be preheated for at least 30

minutes to make them stable. To obtain the highest measurement accuracy, the performance test must be

executed at room temperature or lower; the AC power voltage varies a little only; there must be no abnormal

noise, vibration, dust, or humidity.

Advertising
This manual is related to the following products: