2 output level subsystem – Anritsu MG3641A User Manual

Page 138

Advertising
background image

SECTION 6 GPIB

6-32

6.7.2

Output level subsystem

In the output level subsystem, both “:POWer: SOURce” and “:AMPLitude[:OUT]” are valid for the first and second

layers. (When the first layer is PEWer, the SOURce of the second layer cannot be omitted.) The same processing

is performed in either format. “:AMPLitude[:OUT]” is used in this document.

:AMPLitude[ :OUT]:LEVel <Numeric> <Ampl term> or <Character>

Function

Sets the output level.

Parameter

<Numeric> = –143 dBm to +23 dBm
<Character> = UP

DOWN

Unit

<Ampl term>
(UP,DOWN:<None term>)

Restriction

None

:AMPLitude[ :OUT]:LEVel?

Function

Reads out the output level.

Response

Output level

Restriction

None

:AMPLitude[ :OUT]:LEVel:STEP[ :INCRement] <Numeric> <Ampl term>

Function

Sets the output level step value.

Parameter

<Numeric> = 0.01 dB to 166 dB

Unit

<Ampl term>

Restriction

Unit is dB, only.

:AMPLitude[ :OUT]:LEVel:STEP[ :INCRement]?

Function

Reads out the output level step value.

Response

Output level

Restiction

None

:AMPLitude[ :OUT]:LEVel:UNIT <Ampl term>

Function

Sets the output level unit.

Parameter

None

Unit

<Ampl term>

Restriction

None

:AMPLitude:[ :OUT][ :LEVel]:UNIT?

Function

Reads out the output level unit.

Response

Output level unit = dB

dBm
dBu
V
mV
uV

Restriction

None

Advertising